Details
CBC 5 Control Powerful Refrigerated And Heating Circulator
The CBC 5 control is a powerful refrigerated and heating circulator with strong heating and cooling power for an operating temperature range of - 25 °C to + 200 °C. It is best suited for external applications thanks to its proven pressure and suction pump. High-quality insulation of the device allows for fast heat-up times and reduces heat input at low temperatures. All parameters are manageable (read out, monitored and completely documented) by software (e. g., labworldsoft(r) or NAMUR commands).
-
High performance pressure/suction pump (speed-controlled)
-
Industry standard TFT display for maximum visibility (pump speed, temperature)
-
Adjustable safety circuit
-
Visible and audible alarm
-
Safety drain valve
-
USB / RS 232 interfaces (labworldsoft(r) and enabling online updates of device software)
-
Connection for external temperature probe (PT 100) and optional external solenoid valves (multi-IO port)
-
Safety classification III (FL, DIN 12876)
-
Removable WiCo (wireless controller) enables safe remote operation, e.g. outside of a closed fume hood
Model |
CBC 5 Control |
Appliance type |
Circulation thermostat |
Class designation acc. DIN 12876 |
III |
Identification according to DIN 12876 |
FL |
Cooling agent |
R134a |
Cooling agent quantity |
230 g |
Cooling agent pressure max. |
20 bar |
Heat output |
2500 W |
Cooling capacity (@20°C) |
350 W |
Cooling capacity (@10°C) |
320 W |
Cooling capacity (@0°C) |
270 W |
Cooling capacity (@-10°C) |
190 W |
Cooling capacity (@-20°C) |
80 W |
Cooling capacity (@ -20°C; 3200 rpm) |
130 W |
Cooling capacity (@ 20°C; 3200 rpm) |
400 W |
Working temperature min. |
-25 °C |
Working temperature max. |
200 °C |
Operating temperature min. |
-25 °C |
Temperature display |
yes |
Temperature stability DIN 12876 (@+70°C) |
0.01 ±K |
Temperature control |
PT 100 |
Working temperature sensor |
PT 100 |
Safety temperature sensor |
PT1000 |
Working temperature display |
TFT |
Safety temperature display |
TFT |
Connection for ext. temperature sensor |
PT 100 |
Display resolution |
0.01 K |
Display for operation with ext. sensor |
yes |
Warning function optical |
yes |
Warning function acoustic |
yes |
Warning function excess temperature |
yes |
Warning function insufficient temperature |
yes |
Adjustable safety circuit min. |
0 °C |
Adjustable safety circuit max. |
260 °C |
Sub-level protection |
yes |
Filling volume min. |
5 l |
Filling volume max. |
7 l |
Pump type |
Pressure- / suction pump |
Pump capacity adjustable |
yes |
Pump pressure max. (0 liters discharge flow) |
0.61 bar |
Pressure pump (suction side) (0 liter flow) |
0.45 bar |
Flow rate max. (0 bar back pressure) |
31 l/min |
Pump connection |
M16x1 |
Bath depth min. |
150 mm |
Calibration option |
yes |
Technical data complies with the standard |
DIN 12876 |
Dimensions (W x H x D) |
275 x 690 x 490 mm |
Weight |
40.0 kg |
Permissible ambient temperature |
5 - 32 °C |
Permissible relative humidity |
80 % |
Protection class according to DIN EN 60529 |
IP 21 |
USB interface |
yes |
RS 232 interface |
yes |
Voltage |
230 / 100 - 115 V |
Frequency |
50/60 Hz |
Power input |
3000 W |