Details
FilterCheck 06 : Integrity Test Equipment
- FilterCheck-06 is a 21 CFR compliant, automatic, compact microprocessor controlled integrity test equipment designed for integrity testing of membrane, cartridge and capsule filters.
- This state of the art equipment combines the latest mass flow measurement technology with embedded intelligence systems to provide very high precision and accuracy.
- It also includes additional intelligence in the form of an on board computer with capacitive touch screen and specially developed, user-friendly software with simple menu driven data input.
- This makes the FilterCheck-06 an extremely easy-to-use integrity test system with an added advantage of a large memory to store programs and test results making it ideal for large volume production as well as multi-product processes.
- The Filter Check-06 comes with a special printer with Wi-Fi connectivity for high quality hard copy print outs.
Filter Check Performs:
Features:
Latest Technology for High Precision and Accuracy
Compatibility with Sterile Processing Areas
-
Electropolished SS316L Surface
-
No Instrument Cooling Fan
-
Wireless and Paperless to minimize particlegeneration in sterile areas
Easy Portability
All types of Integrity tests
21 CFR part 11 Compliance for Data Reliability and Security
Self Diagnosis
Shows Real Time Test Status
Large Storage Space >10GB for
-
Test Templates
-
Test Reports
Inbuilt Equipment Safety with separate Hydrophobic Filters for
Specifications:
FEATURES |
Catalogue No |
LSPF01 |
Stainless steel body |
SS316L |
Dimension (Main test Unit) |
35 x 32 x 12 cm |
Weight |
9Kg |
Voltage |
100-240V AC |
Input Frequency |
50Hz |
Power |
75W |
Battery Pack |
14.4 V, 4.5 AH |
Capacitive Touchscreen |
7" |
Fuse |
4A |
Battery Charger (Inbuilt) |
Yes |
Power Backup |
3 hours |
External Wireless Printer |
Yes |
Charging time |
2 hours |
OPERATING SYSTEM |
Android |
Jelly Bean 4.2.2 |
OPERATING RANGE |
Maximum Inlet Pressure |
100 psi |
Ambient Temperature |
15° - 35°C |
Relative Humidity |
10 - 80% |
MEASURING RANGE |
Diffusion Flow |
0.2 - 100 ml/min |
Bubble Point |
15 - 80 psi |
Water Intrusion Rate |
0.03 - 10 ml/min |
ACCURACY |
Diffusion Flow |
+ 5% or 1ml whichever is greater |
Bubble point |
+ 0.7 psi |
Water Intrusion Rate |
+ 5% or 0.02ml whichever is greater |
SUPPORT DOCUMENTS |
Certificate of Quality |
Yes |
Calibration Report |
Yes |
Installation Qualification |
Yes |
OQ (Test Equipment) |
Yes |
OQ (Application Software) |
Yes |
Performance Qualification |
Yes |