100mA Radiographic X-ray Machine JYSX0101
USE
General radiography, Filter-ray-filter radiograph
Unit Configuration
Control, HT generator unit, Ht cable, X-ray tube unit, Radiographic table-X-ray source column, Collimator
Function
The unit can used for general radiography, Bucky radiography of body’s head, chest, foot and hand, bones in lying
Feature
1. meter-indicating
2. single-table single tube
3. single-phase, full wave rectification
4. stationary anode x-ray tube assembly
5. X-ray tube filament s voltage regulator, space charge compensator
6. Radiograph plain table wimple structure and easy assembling
MAIN TECHNICAL INDEXES
ITEM |
CONTENTS |
INDEX |
Demand of power supply |
Capacity |
Two-phase 30KVA |
Voltage Frequency |
380V/220V ±1050HZ ±0.5Hz |
Intemal resistance |
2Ω(380V) 0.6 Ω(220V) |
power supply |
Adjustable range |
380±10% at220V being adjustable continuously |
Radiograph |
Voltage |
50-90KV |
current |
Small focus: 15mA,Big focus:30mA 60mA 100mA |
time |
0.04~6.35s 23grades together grading a according to R10 coefficient |
High-voltage generator |
Capacity |
7KVA |
Max DC output voltage |
90KV |
Max DC output current |
100mA |
X-Ray tube |
Model |
0XD4-2.9/100 |
Focus |
Small focus 1.8×1.8mm big focus4.3×4.3mm |
Radiographic |
Length×width×height of bed surface |
2000mm×665mm×700mm |
|
Moving range of bed surface |
640mmin Length,200mm in breadth |
table |
Ray-filter of table |
Travel in length ≥50mm |
Grid density N=28 |
Convergence distance f.=100cm |
Grid ratio r=8 |
Column of X-ray tube unit |
Moving Longitudinally along |
1000mmor 1800mm |
Radiographic table |
Distance to friend as moving up and down (along column) |
650~1800mm |
Rotary around the center of cross arm |
±180° |
Rotary around the shaft line of X-ray annular tubes |
-10°~60°~ +120° |
Max size of cassette |
356mm×432mm(14×17) |
Net Weight: 600kg Gross: 695kg Packing Size: 230*112*111