Details
Fetal Doppler
Product Specification
Safety: Complies with: IEC 60601-1:1988+A1:1991+A2:1995
Classification:
-
Anti-electroshock Type: Internally powered equipment.
-
Anti-electroshock Degree: Type B applied part
-
Harmful Liquid Proof Degree:
-
Main unit :Ordinary equipment (sealed equipment without liquid proof).
-
Probe: Prevent from water splashing, degree of protection: IPX4.
-
Degree of Safety in Presence of Flammable Gases: Equipment not suitable for use in
-
presence of flammable gases.
-
Working System: Continuous running equipment.
-
EMC: Group I Class B.
-
Suitable Using Range: Suitable for use after the 12th week of pregnancy.
Physical Characteristic
Environment Working:
Transport and Storage:
FHR Performance
-
FHR Measuring Range: 50~240BPM (BPM: beat per minute)
-
Resolution: 1BPM
-
Accuracy: ±2BPM
-
Power consumption:<0.8W
-
Auto Shut-OFF: After 1 minute no signal, power off automatically.
-
Battery Type Recommended: Two pieces of 1.5 V DC battery (SIZE AA LR6).
Probe:
-
Nominal Frequency: 2.0MHz
-
Working Frequency: 2.0MHz±10%
-
P-: <0.5MPa
-
Iob: <10 mW/cm2
-
Ispta: <50mW/cm2
-
Ultrasonic Output Intensity: Isata<5 mW/cm2
-
Ultrasonic Output Power: P<10 mW
-
Working Mode: Continuous wave Doppler
-
Effective Radiating Area of Transducer: 157mm2±15%