- Description
-
Details
Medium Sterilizer
This series of pulsation vacuum sterilizer uses saturated steam as its sterilization medium, because the saturated steam releases a large amount of latent heated physical properties in the stage of condensation. Applicable for high temperature sterilization equipment to be used in medium products. It has been widely used in pharmaceutical, medical, scientific research projects.
Main Parameters and UtilitiesDesign Pressure 0.245 MPa Design Temperature 1390 C Vacuum -0.09 MPa Water Source Pressure 0.15-0.3 MPa Steam Pressure 0.3-0.7 MPa Working Pressure 0.225 MPa Working Temperature 105-1340C Temperature Equilibrium ≤± 10C Compressed Air Pressure 0.3-0.7 MPa Power AC 380V; 50 Hz Specifications & Ordering Details
Catalogue No Overall Size
L x W x H (mm)
Chamber Size
L x φ
Steam
Consumption (Kg)
Tap Water
Consumption (KG)
Power
(KW)
Net Weight
(KG)
JWG-P06 1500 X 1360 X 1950 1200 X 610 X 910 30 45 3 + 1400 JWG-P08 1800 X 1360 X 1950 1500 X 610 X 910 40 50 3 1600 JWG-P12 1750 X 1440 X 1950 1450 X 680 X 1180 48 55 4 1800 JWG-P15 2150 X 1440 X 1950 1850 X 680 X 1180 55 60 4 2000 JWG-P20 1950 X 1750 X 2200 1600 X 900 X 1400 65 65 4 2500 JWG-P25 2300 X 1750 X 2200 1950 X 900 X 1400 75 75 6 3000 JWG-P30 2700 X 1750 X 2200 2400 X 900 X 1400 90 90 6 3500 JWG-P35 2700 X 1750 X 2200 2400 X 900 X 1400 105 105 6 4000 JWG-P40 2700 X 1750 X 2200 2400 X 900 X 1400 120 120 8 4500 JWG-P50 2700 X 1750 X 2200 2400 X 900 X 1400 135 135 8 5000 Structure and Characteristic of The Equipment
Body of the sterilizer:
- Inspection procedures.
- Tank of the horizontal rectangular structure is made from S30408.
- Door sealing: Pneumatic sealing; both doors are interlocked and completely conform to the requirements of the specification.
- The equipment has standard GMP authentication interface.
Sterilizer piping system:
The pipeline is designed with the best configuration. By employing imported parts and quality domestic brands into main components.Sterilizer control system: with the adoption of Siemens PLC plus Siemens touch screen, this system is easy to implement the program selection, parameter setting, equipment operation, report processing and other functions. It has perfect safety measures. The F0 value and the temperature time gives double guarantee for sterilization. A single parameter control is also available. It has a perfect sterilization records.it is designed and manufactured in accordance with GB150 "the Pressure Vessel”, and TSG R0004-2009 the “Safety and Technology Specification for Fixed Monitor Pressure Vessels” and related
- Additional Information
-
Additional Information
Code No - Reviews
-