RST Spindles
- The choice of coaxial cylinder, cone and vane spindles accommodates viscosity measurement of a wide variety of materials.
- The RST series Rheometers are available as three different models each with a different spindle geometry:
- RST-CC Rheometers use coaxial cylinder spindle geometry
- RST-CPS Rheometers use cone and plate spindle geometry
- RST-SST Rheometers use vane spindle geometr
SPECIFICATIONS
RST Spindle |
RS Rheometers with serial numbers beginning with “302” and “303” use different spindles. Call for details. |
SPINDLE |
VISCOSITY RANGE |
SHEAR RATE |
MAX. SHEAR STRESS |
SAMPLE VOLUME |
COAXIAL |
Pa•s |
sec-1 |
Pa |
mL |
CCT-DG |
0.00005-4.07K |
0.043-5.64K |
177 |
15.7 |
CCT-40 |
0.0003-27.6K |
0.0215-2.79K |
594 |
68.5 |
CCT-25 |
0.002-177K |
0.013-1.67K |
2.28K |
16.8 |
CCT-14 |
0.012-1M |
0.013-1.68K |
13K |
3.4 |
CCT-8 |
0.065-5.41M |
0.013-1.672K |
69.6K |
1.0 |
CONE |
Pa•s |
sec-1 |
Pa |
mL |
RCT-25-1 |
0.005-407K |
0.06-7.8K |
24.4K |
0.1 |
RCT-25-2 |
0.01-814K |
0.03-3.9K |
24.4K |
0.2 |
RCT-50-1 |
0.0006-50.9K |
0.06-7.8K |
3.05K |
1.0 |
RCT-50-2 |
0.0012-101K |
0.03-3.9K |
3.05K |
2.0 |
RCT-75-1* |
0.0002-15K |
0.06-7.8K |
905 |
2.5 |
RCT-75-2* |
0.0004-30K |
0.03-3.9K |
905 |
5.0 |
PLATE |
Pa•s |
sec-1 |
Pa |
mL |
RPT-25 |
0.03-2.49M |
0.013-1.7K |
32.6K |
0.5 |
RPT-50 |
0.002-155K |
0.027-3.4K |
4.07K |
2.0 |
RPT-75* |
0.0004-30.7K |
0.04-5.1K |
1.2K |
4.5 |
VANE SPINDLE |
VANE LENGTH (mm) |
VANE DIAMETER (mm) |
SHEAR STRESS (Pa) |
VT-10-5 |
10 |
5 |
330-210K |
VT-20-10 |
20 |
10 |
41-27K |
VT-20-20 |
20 |
20 |
9-5.9K |
VT-30-15 |
30 |
15 |
12-8K |
VT-40-20 |
40 |
20 |
5.2-3.4K |
VT-40-40 |
40 |
40 |
1.2-740 |
VT-50-25 |
50 |
25 |
2.7-1.7K |
VT-60-8 |
60 |
8 |
24-15K |
VT-60-15 |
60 |
15 |
7-4.3K |
VT-60-30 |
60 |
30 |
1.6-1K |
VT-80-40 |
80 |
40 |
0.7-420 |
VT-80-70 |
80 |
70 |
0.2-120 |
*For use with water bath version only. 1 Pa · s = 1,000 cP K = 1 thousand M = 1 million
Note: 1. Values based on minimum speed of 1 RPM and maximum speed of 1000 RPM
2. 75 mm plates cannot be used with Peltier Plate or electrically heated rheometers