Details
Two Probe Method With USB Interface
For resistivity measurement of insulators at different temperatures
The Two Probe Method is one of the standard and most commonly used method for the measurement of resistivity of very high resistivity samples – near insulators. The resistivity measurement of such samples is beyond the range of Four Probe Method.
DESCRIPTION OF THE EXPERIMENTAL SET-UP
1. Two Probes Arrangement, LSTPA-01
It has two spring load contact probes. These probes move in a pipe and are insulated by Teflon washers. This probes arrangement is mounted in a suitable stand, which also holds the sample plate and RTD sensor.
The stand also serves as the lid of PID Controlled Oven. Teflon coated leads are provided for connecting wit High Voltage Power Supply EHT-11C and Digital Picoammeter LSDPM-111C. With this set-up assuming max. Voltage = 1500V; current 10x10-12 A (max) and thicknexx of sample 1mm, the resistivity of the sample could be measured upto1014 ohm.cm..
2. PID Controlled Oven, LSPID-TZ
The unit is a high quality PID controller wherein the temperatures can be set and controlled easily. The P, I and D parameters are factory set for immediate use however the user may adjust these for specific applications as well as auto-tune the oven whenever required.
3. LSDPM-111C
4. LSEHT-11C
Ordering Information: