No pulse delivery : full uniform motion , smooth pulseless ;
-
Without cleaning : fluid
-
Transport different characteristics just replace the syringe ;
-
Adaptability : suitable for a variety of different brands of syringes ;
-
High precision control : control error ± 3%.
-
Comprehensive index -type injection pump Products in the following table :
-
Injection rate 0.1m1 / h -500.0 ml / h, step : 0.1 ml / h Applicable syringe 5 ml, 10 ml, 20 ml, 30 ml, 50 ml
-
Speed setting range 5 ml: 0.1 - 50.0 ml / h, 10 ml: 0.1 - 100.0 ml /h, 20 ml: 0.1 - 200.0 ml / h, 30 ml: 0.1 -300.0 ml / h, 50 ml: 0.1 - 500.0 ml / h
-
Can be synchronized to display the cumulative increase in the amount of time the cumulative injection volume display 0.1-999.9 ml, rapid injection of state
-
Rapid injection speed 5 ml: 50.0 ml / h, 10 ml: 100.0 ml / h, 20 ml: 200.0 ml / h, 30 ml: 100.0 ml / h, 50 ml: 500.0 ml / h
-
Injection Accuracy ± 3% ( after trimming setting up to ± 2%)
-
KVO rate of 1m1 / h, keeping the vein open state Power Requirements AC 85V - 165V, 50 HZ
-
Built- in rechargeable lithium-ion battery battery, DC8.4V, 1600mAh; After fully charged 25.0 ml / h continuous work more than 5 hours
-
The flow rate of the display information , the amount of accumulation , syringe size , battery capacity, battery charging , AC power source, complete the display , the display clogging
-
Alarm alarm function is completed , nearing completion alarm, clogging alarm, low battery capacity, syringe abnormal , no AC alarm
-
Outline maximum size 303 . 126 . 118 mm (LxWxH)
-
Maximum weight is less than 2kg