jainsonsindia partners
 Enquiry Cart

Horizontal Cylindrical Pressure Steam Sterilizer

More Views

Horizontal Cylindrical Pressure Steam Sterilizer

Code:
Horizontal Cylindrical Pressure Steam Sterilizer 
(Microcomputer Control)
 Characteristics: 
1.Sterilizing course automatic sterilization controlled by computer, easy to operate.
2. With drying function, suitable dressing drying.
3. With over-temperature, over-pressure auto-protect device.
4. The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can’t be start on if the doesn’t closed well.
5. The safety value will be open automatically when the inner pressure over 0.24MPa and the steam is exhausting to the water tank.
6. The power will be cut-off automatically, cut-off the water and alarm when the machine lacking water.
7. The chamber of the sterilizer is made of stainless steel.
8. With printer.

Technical data
MODEL JWS-150YDD JWS-200YDD JWS-280YDD
Sterilizing chambervolume 150Lφ440×1000mm 200Lφ515×1000mm 280Lφ600×1000mm
working pressure 0.22 MPa
working  temperature 134℃
Adjustment of temperature 40℃ - 134℃
Time of sterilization 0-99 min
Time of drying 0-99 min
Heat average ≤± 2℃
Power 9KW/380V 50Hz 9KW/380V 50Hz 12KW/380V 50Hz
Dimension 1400×650×1600 1400×750×1700 1520×910×1900
Transport dimension 1560×820×1820 1560×910×1880 1680×1080×2100
G.W/N.W 430/340 KG 436/350 KG 570/462 KG
 
Description

Details

Horizontal Cylindrical Pressure Steam Sterilizer 
(Microcomputer Control)
 Characteristics: 
1.Sterilizing course automatic sterilization controlled by computer, easy to operate.
2. With drying function, suitable dressing drying.
3. With over-temperature, over-pressure auto-protect device.
4. The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can’t be start on if the doesn’t closed well.
5. The safety value will be open automatically when the inner pressure over 0.24MPa and the steam is exhausting to the water tank.
6. The power will be cut-off automatically, cut-off the water and alarm when the machine lacking water.
7. The chamber of the sterilizer is made of stainless steel.
8. With printer.

Technical data
MODEL JWS-150YDD JWS-200YDD JWS-280YDD
Sterilizing chambervolume 150Lφ440×1000mm 200Lφ515×1000mm 280Lφ600×1000mm
working pressure 0.22 MPa
working  temperature 134℃
Adjustment of temperature 40℃ - 134℃
Time of sterilization 0-99 min
Time of drying 0-99 min
Heat average ≤± 2℃
Power 9KW/380V 50Hz 9KW/380V 50Hz 12KW/380V 50Hz
Dimension 1400×650×1600 1400×750×1700 1520×910×1900
Transport dimension 1560×820×1820 1560×910×1880 1680×1080×2100
G.W/N.W 430/340 KG 436/350 KG 570/462 KG
 
Additional Information

Additional Information

Code No
Reviews