Details
Rapid Cooling Sterilizer
This series of rapid cooling sterilizer applies saturated pure steam as its sterilization medium. The saturated pure steam releases a large amount of latent heated physical properties in the stage of condensation, which keeps the items to be sterilized under a status of high temperature and humidity. After a period of insulation in such status, the
Sterilization can be achieved. It features with rapid and high sterilization effects while working together with the rapid condensing unit.
Applicable for equipment and devices requiring sterilization and cooling functions for various large glass bottles and soft bag infusion products with moisture and heat resistance properties. It has been widely used in pharmaceutical, medical, and scientific research projects.
Main Parameters and Utilities
Design Pressure |
0.245 MPa |
Design Temperature |
1390 C |
Vacuum |
-0.09 MPa |
Water Source Pressure |
0.15-0.3 MPa |
Steam Pressure |
0.3-0.7 MPa |
Working Pressure |
0.225 MPa |
Working Temperature |
105-1340C |
Temperature Equilibrium |
≤± 10C |
Compressed Air Pressure |
0.3-0.7 MPa |
Power |
AC 380V; 50 Hz |
Specifications & Ordering Details
Catalogue No |
Overall Size L x W x H (mm) |
Chamber Size L x W x H (mm) |
Steam Consumption (Kg) |
Net Weight (KG) |
JWG-K12 |
1750 X 1650 X 1950 |
1450 X 680 X 1180 |
70 |
1650 |
JWG-K15 |
2150 X 1650 X 1950 |
1850 X 680 X 1180 |
85 |
1900 |
JWG-K20 |
1950 X 2000 X 2200 |
1600 X 900 X 1400 |
100 |
2550 |
JWG-K25 |
2300 X 2000 X 2200 |
1950 X 900 X 1400 |
140 |
2950 |
JWG-K30 |
2750 X 2000 X 2200 |
2400 X 900 X 1400 |
180 |
3400 |
Structure and Characteristic of The Equipment
Body of the sterilizer:
-
It is designed and manufactured in accordance with GB150 "the Pressure Vessel", and TSG R0004-2009 "Safety and Technology Specification for Fixed Monitor Vessels and related inspection procedures.
-
Tank of the horizontal rectangular structure is made from S30408.
-
Door sealing: Pneumatic sealing; both doors are interlocked and completely conform to the requirements of the specification.
-
The equipment has standard GMP authentication interface.
Sterilizer piping system:
-
The pipeline is designed with the best configuration. By employing imported parts and quality domestic brands into main components.
-
Sterilizer control system: with the adoption of Siemens PLC plus Siemens touch screen, this system is easy to implement the program selection, parameter setting, equipment operation, report processing and other functions. It has perfect safety measures. The F0 value and the temperature time gives double guarantee for sterilization. A single parameter control is also available. It has a perfect sterilization record
Load Capacity of The Equipment
Catalogue No |
50 ml |
100 ml |
250 ml |
500 ml |
JWG-K12 |
3040 |
1956 |
920 |
496 |
JWG-K15 |
3960 |
2520 |
1200 |
652 |
JWG-K20 |
5920 |
3570 |
1650 |
1105 |
JWG-K25 |
7200 |
4515 |
2046 |
1420 |
JWG-K30 |
8960 |
5460 |
2508 |
1680 |