MNRU2101B Fixed Cradle + 3 Drawers
Code:
MNRU2101B Fixed Cradle + 3 Drawers
General Description
- Mobile radiant heat warmer-fixed height
- Mounted on 4 swivel castors (100mm) - 2 castors with brake
- Baby tray with mattress along with head up/down facility and X-ray cassette holder
- Mattress density approx. 21-25 kg/m3 and removable washable water proof cover
- 6mm thick - Transparent acrylic drop down and lockable side panel
- Side tray for monitor and infusion rod for syringe pump/SS IV rod
- 2 shelves below baby bed for under surface Phototherapy and storage
- Metallic, swivelling hood suspended above the table integrates heating element and overhead light
- Heating element – Ceramic heater (650 W) with parabolic reflector, protected by metallic grid
Oxygenation:
- Integrated support for 10L oxygen Cylinder
- Oxygen regulator with pressure gauge, flow meter and humidifier bottle along with silicone tubing for delivery of oxygen from humidifier bottle to patient
Suction:
- Integrated Slow Suction unit with Vacuum Regulator, Vacuum Gauge and Autoclavable
- Fluid Jar 500ml (2 nos.) and Silicone tubing
Resuscitation:
- Silicone manual Resuscitator size Infant (250ml) with 00 and 01 size face masks.
- Laryngoscope - Penlight handle with size 0 and 1 straight blade
- Silicone Airways size 000, 00 and 0
Temperature Controller B
- Big red LED display for skin temperature, set temperature & timer
- LED indicator for selected mode & alarm
- Bar-graph for heater power
- Integrated timer – 1 to 09 min with count-up feature
- Temperature range – skin: 34 to 38°C (user pre-settable)
- Monitoring of skin temperature by means of sensor, range: 30 to 42°C
- Control unit – auto over-temperature cut-off with audio-visual alarms according to pre-set temperature to avoiding overheating
- Display indicates system errors, sensor failure, alarm for high/low, over/under temperature for skin, heater failure, power control device failure.
- Audio-Visual alarm for power failure