Hydrolysis Unit - HU 6
Code:
Hydrolysis Unit - HU 6
Optimum solution for the acid/basic hydrolysis of food and feed samples prior to solvent extraction for total fat analysis.
Improves Analysis
- Hydrolysis / digestion by acid or base
- Ensures conformity with AOAC standard
- No sample transfer – no sample loss
- Crucibles go directly into SER 148 or SER 158
- Reduces handling to a minimum
- Automatic extraction of unwanted fraction
- No exposure to vapors
Recommendations
- Select a glassware kit and accessories to match workload
- Hydrolysis is carried out with hydrochloric acid for approximately 60 minutes at 170 °C.
The hydrolyzed sample is then filtered in a glass crucible and washed with warm deionized water in order to eliminate the residues of hydrochloric acid
Technical Specifications
GENERAL FEATURES |
Construction material: |
epoxy painted stainless steel structure |
Number of samples: |
6 samples |
Set temperature: |
display visualization |
Countdown: |
display visualization |
Display: |
LCD |
Number of programs: |
20 |
Selectable language: |
I, F, UK, E, D, T |
Power: |
1350 W |
Dimensions (WxHxD): |
355x590x450 mm (14.0x23.2x17.7 in) |
Weight: |
14.5 Kg / 32.0 lb |
|
|
PERFORMANCE |
Temperature range: |
from room temp. to 200 °C |
Temperature calibration: |
automatic |
Stability of the heating block temperature: |
± 0,5°C |
Homogeneity of the heating block temperature: |
± 0,5°C |
Precision of the heating block temperature: |
± 0,5°C |
Reaching of set temperature: |
acoustic and visual signal |
Including Accessories:
- EDPM Tube Ø 6.4 x 11.2mm
- Celite 1 kg
- Glass Sand 2 kg
Ordering Information:
Catalogue No. |
Description |
Power Supply |
F30300110 |
HU 6 |
230V / 50-60Hz |
F30310110 |
HU 6 |
115V / 50-60Hz |