- Description
-
Details
Sath Type Leakage Detection Sterilizer
This series high temperature pulsation vacuum sterilizer applies deionised water as its sterilization medium. With the internal high temperature deionised water circulation, it realizes sterilization by evenly spraying the processed deionised water onto items to be sterilized for a period. With a wide temperature range and uniformity, this equipment has successfully prevented the secondary pollution during the procedure. It ensures the scrap rate up to 100% by implementing means of vacuum and colour leakage detection methods. Applicable for equipment and devices requiring high temperature sterilization, leakage detection and cleaning processing for injection and oral liquid products contained in ampoule, vials and tube-type bottles. It has been widely used in pharmaceutical, medical, and scientific research projects.
Main Parameters and UtilitiesDesign Pressure 0.245 MPa Design Temperature 1390 C Vacuum -0.09 MPa Water Source Pressure 0.15-0.3 MPa Steam Pressure 0.3-0.7 MPa Working Pressure 0.225 MPa Working Temperature 80-1340C Temperature Equilibrium ≤± 10C Compressed Air Pressure 0.3-0.7 MPa Power AC 380V; 50 Hz Specifications & Ordering DetailsCatalogue No Overall Size
L x W x H (mm)
Chamber Size
L x W x H (mm)
Steam Consumption (Kg) Consumption Of
Cleaning Water (KG)
Consumption Of
Deionised Water (KG)
Sterilization Vehicle Net Weight (KG) JWG-SS12 1750 X 1650 X 1950 1450 X 680 X 1180 85 760 140 2 1750 JWG-SS15 2150 X 1650 X 1950 1850 X 680 X 1180 100 1000 170 2 2000 JWG-SS20 1950 X 2000 X 2200 1600 X 900 X 1400 140 1250 220 2 2750 JWG-SS25 2300 X 2000 X 2200 1950 X 900 X 1400 180 1500 260 2 3150 JWG-SS30 2750 X 2000 X 2200 2400 X 900 X 1400 240 1800 300 3 3550 JWG-SS40 3550 X 2000 X 2200 3200 X 900 X 1400 300 2100 420 3 4050 Structure and Characteristic of The EquipmentBody of the sterilizer:-
It is designed and manufactured in accordance with GB150 "the Pressure Vessel", and TSG R0004-2009 Safety and Technology Specification for Fixed Monitor Vessels and related inspection procedures. -
Tank of the horizontal rectangular double-deck structure is made from S30408. -
Door sealing: Pneumatic sealing; both doors are interlocked and completely conform to the requirements of the specification. -
The equipment has standard GMP authentication interface.
Sterilizer piping system:The pipeline is designed with the best configuration. By employing imported parts and quality domestic brands into main components, this system presents online sterilization option.Sterilizer control system:with the adoption of Siemens PLC plus Siemens touch screen, this system is easy to implement the program selection, parameter setting, equipment operation, report processing and other functions. It has perfect safety measures. The F0 value and the temperature time give double guarantee for sterilization. A single parameter control is also available. It has a perfect sterilization records.Load Capacity of The EquipmentCatalogue No 1 ml 2 ml 5 ml 10 ml 20 ml JWG-SS12 103008 71808 29808 17360 10608 JWG-SS15 133632 92048 38880 22568 13014 JWG-SS20 184704 123760 48576 33600 19008 JWG-SS25 255744 169344 66792 45600 25344 JWG-SS30 291264 196560 78936 52800 30144 JWG-SS40 415584 278640 109296 75600 40464 -